Product
Polypropylene Hybrid Laminar Airflow
TopAir’s Laminar Airflow Ductless Fume Hoods
combine the functionality of ductless fume hoods and
laminar clean benches.
The product is equipped with dual filtration system:
1. Supply system with a pre-filter, HEPA filter and a
fan that creates laminar down-flow towards the
working area.
2. Exhaust system with carbon filter and fan that
collects the down-flow and creates protective
inflow on the front sash opening. Both inflow and
down flow pass through the carbon filter before
the air is exhausted back into the lab.
Structure :
• 8 mm welded polypropylene with high chemical
resistance
• Double back wall with baffle
• Built-in sealed polypropylene or optional stainless
steel worktop
• Tempered glass, full-opening sliding front sash with
counterweight
• Extraction collar for external exhaust
• 10.1” color touchscreen display
• VAV with air flow alarm and display
• Automatic laminar air velocity, manual adjustment for inflow includes a safety shutdown
system for the down-flow when inflow system
fails
• VOC sensor and alarm
• Chamber alert light (red)
• Cleanliness level: Class 100/ISO 5
• 800 lux LED lighting separated from the work
area
• Includes polypropylene lower base cabinet
• CE certified